OCTOPUS 900
The Most Versatile 90° Full Field Perimeter for Static- and Kinetic Perimetry
It incorporates standard White-on-White and Blue-on-Yellow static perimetry.
With an optional software program, the OCTOPUS 900 can be upgraded to perform true Goldmann kinetic perimetry.
The OCTOPUS 900 covers all the needs in clinics and in scientific applications without limitation.
Reliable Precision
Speed and Ease of Use
Patient Comfort
Goldmann Kinetic Perimetry*
OCTOPUS 900 Technical specifications
Optical specifications
Optical System
Spherical Goldmann Cupola
Usable temporal range
Horizontal +89°/-89°, Vertical +60°/-67°
Stimulus generation
Mirror projection system
Resolution
0,1°
Stimulus duration
100, 200, 500, 1000, infinite
Dynamic range
47dB (max. 1910cd/m²)
Stimulus colors
White, Blue* (440nm), Red* (610nm)
Stimulus size
Goldmann I, II, III, IV, V
Background
1,27cd/m² / 10cd/m² (white), 100cd/m² (yellow )*
Reliability measures
Fixation
Realtime blink and fixation control (60 frames/second)
Eye position
Automated Eye Tracking*,**
Lensholder
Automated positioning and distance adjustment**
Head position
Non contact head sensor
Test speed
Adaptive test speed based on patient's reaction time
Test methods
White/White
Standard automated perimetry
Blue/Yellow
Short wavelength automated perimetry*
Red/White
Color perimetry*,**
Kinetic
Manual and Automated Goldmann Kinetic perimetry*
Flicker
CFF Critical Flicker Fusion Frequency perimetry*,**
Threshold & Screening test strategies
Normal strategy
8-12 min (4-2-1 bracketing strategy)
Dynamic strategy
5-8 min (adaptive step size)
TOP*
2-3 min (Tendency Oriented Perimetry)*
LVS
Low vision (Bottom up) Goldmann V strategy
2-LT
3 zone (normal, relative, absolute defect)
1-LT
2 level (seen, non seen)
Multi-phase strategy
Consecutive threshold testing of relative defects and normal areas
Program library
M
Physiological macular test 5°/ 12°/ 30°
G
Physiological general test 30°/ 60°
32
Equidistant 30° test
07
Cortically based overview test to 75°
LVC, LVP
Low vision programs for end stage follow up
D1, N1
Programs for Diabetes, Neuro-ophthalmology
BT, BG, FG, Esterman
General legal tests**
Country dependent
Country dependent static and kinetic expert opinion tests**
Custom Test*
Individual test programming with a graphical user interface*,**
Octopus features
Fixation control
Real time infrared eye control automatically repeating stimuli that interfere with blinking and delaying stimuli presentation during non fixation
Pupil measure
Automatic averaging pupil size measurement
Easy follow up
Follow up button ensuring identical test parameters**
Flexible test procedures
Screening programs can be used for threshold testingCentral programs can be extended to the periphery
Procedure codes
Standard operating procedures with all parameters and printout definitions can be automated
Reaction time
Adaptive interstimulus time for static perimetry, patented correction for reaction time in kinetic perimetry*
Electrical and mechanical specifications
Power requirements:
100 – 120VAC / 220 - 240 VAC 50/60Hz
Power consumption:
145VA / 165 VA
Measures (W x L x H):
648mm x 519mm x 796 mm
Weight:
25 kg
Temperature:
Operation: +15°C ... +40°C
Light source lifespan
> 20'000 hours for stimulus, background and fixation
Interface
Ethernet T-100
Classification:
Class I, Type B
Conformity
CE: 93/42/EEC, 89/336/EEC, and conformity module ACSA C & US
Compliance
IEC / EN60601-1:2005, EN 60601-1-2: 2001 + prA1, IEC 60601-1-2: 2001 + A1: 2004, EN ISO 15004-1: 2006, EN ISO 12866: 1999 + AC:2000, IEC 60825-1: 2001-08 Ed. 1.2, CIE S 009/E: 2002
Specifications are subject to change without notice
* Pro version (available as options for the Octopus 900 Basic)