O'300 PRO PERIMETER

O'300 PRO PERIMETER INCLUDES TOP STRATEGY, BLUE/YELLOW TESTING, LAN INTERFACE BOARD,FLICKER PERIMETRY,CCFF TESTING,AUTOMATIC EYE TESTING, ALL PRINTOUT FORMATS, CUSTOM TEST PROGRAMS, EYE SUITE LITE ADVANCE PRO PERIMETRY SOFTWARE.

In design and performance, the OCTOPUS 300 Series presents the state-of-the-art compact perimeter. In particular, with its free rotating optical system, the perimeter can be operated from all sides, to enable its use in the tightest space and in ambient light.

Reliable Precision

  • 100% eye fixation control – if fixation is lost, no stimulus is presented.
  • Diagnostic programs with test stages prioritising the most crucial field areas.
  • With the unique Eye Tracking feature (option), the perimeter follows automatically the eye movements.
  • Fast software for significantly shorter examinations avoid errors caused by patient and retinal fatigue.
  • Continuous measurement of the pupil size allowing accurate assessment and comparison of data.

Speed and Ease of Use

  • The free rotating examination unit permits operation from all sides, to enable the use in tight spaces and in ambient light.
  • Simple operation on a color input screen guided by a dialog in any language.
  • With the optional Tendency Oriented Perimetry (TOP) test strategy, a full threshold examination takes less than 2:30 min.
  • Electronic eye fixation adjustment without disturbing the patient.
  • The clear Seven-in-One print-out presents the most complete information needed for the assessment of the results.
  • During the examination, a real time display of statistical data provides on-line information about the status of the patient’s visual field. Also, the test procedure can be interrupted at any time and continued later without losing data.
  • A built-in training program facilitates giving instructions and shortens the learning process.
  • Software is easily updated to the latest version by the exchange of a software module or by downloading from the Internet.

Patient Comfort

  • The ergonomic design maximizes patient comfort.
  • Faster testing results in better patient cooperation.
  • Cold light sources avoid heat dissipation and eliminate the noise of cooling fans.
  • Patients are encouraged to blink regularly without the risk of introducing erroneous results


 

 

OCTOPUS 301

OCTOPUS 311

Optical system

30° direct projection

x

x

Installation

No dark room conditions required

x

x

Design and access

Free rotating examination unit, access from all sides

x

x

Operation

Color LCD input screen

x

x

Work log counter

Ongoing counting of examinations

x

x

Dialog language

All languages available in Windows®

x

x

Eye monitor

Infrared sensitive eye camera and video display

x

x

Eye fixation control

Electronic eye fixation control

x

x

Automatic fixation adjustment

Perimeter automatically adjusts for fixation loss

option

x

Pupil size measurement

Continuous average pupil size measurement

x

x

Examination programs

Programs G1, 32, M2, ST and LVC

x

x

Test strategies

Normal 4-2-1dB, Low vision, Dynamic and 2LT (Qualitative 3-zone) strategy

x

x

TOP strategy

option

option

Automatic procedures

Automatic test settings available for all programs

x

x

Stimulus sizes

Goldmann III and V

x

x

Stimulus intensity

0-4’800asb

x

x

Background illumination

31.4 / 314asb

x

x

Printer output

USB output for connection to standard inkjet printers

x

x

Print-out formats

Selection of 8 print-out formats, including Two-on-One format

x

x

Data output to PC

Data transfer in conjunction with PeriTrend (option)

x

x

Software upgrades

Exchange of software module or transfer of updates from PC by CD or downloading from Internet

x

x

User defined programs

5 custom tests and free selection of catch trials

option

x

Special Perimetry

Blue-on-Yellow Perimetry

-

x

Flicker Perimetry

option

x

Network connection

Integrated LAN adapter with Ethernet interface
Input of patient data from an external software system

option

x

Remote control

PC control of perimeter functions

x

x

Power requirements

100-240VAC, 40VA